September 17, 2026

Written by Austin Tyner

Regulatory Review by Neel Parmar

In the submission of medical devices, the key to a favorable marketing authorization decision is sufficient evidence. When applications lack that evidence, the FDA can identify where that evidence, or the explanation behind it, falls short through a Deficiency Letter. During substantive review, deficiency letters typically cite a gap between what evidence the applicant has presented and what evidence the FDA needs. Some gaps can be addressed with clarification and context, but many cannot be closed without new testing to meet the evidentiary burden. This article addresses the medical device deficiency letter as part of the broader review sequence for a 510(k), De Novo request, or PMA, and gives the applicant a practical, step-by-step approach for building a response.

The Deficiency Letter in the FDA Review Sequence

FDA device reviews move through a defined sequence of stages, each with its own point at which the agency can determine that something necessary is missing. For a 510(k) or De Novo request, that sequence runs from acceptance review, to substantive review, to a final decision. A PMA follows a comparable sequence but includes both acceptance and filing reviews before substantive review begins.

A deficiency letter is not the only correspondence FDA can issue when they find application insufficiencies, and it is important to note that it is not interchangeable with the letters issued earlier in the sequence. The table below describes these correspondences and the criteria that trigger them.

Output Review Stage Trigger
Refuse to Accept (RTA) Acceptance Review Submission is missing elements required for administrative completeness before substantive review can begin.
Not Filed Filing Review (PMA only) PMA lacks the information necessary for FDA to begin a substantive review.
Interactive Review Inquiry Substantive Review Reviewer needs a discrete piece of information that can generally be resolved without placing the submission on hold.
AI Request / Major Deficiency Letter Substantive Review Formal request for information FDA needs to reach a decision; places the submission on hold until a complete response is received.

What to Expect in a Deficiency

FDA guidance directs reviewers to build each deficiency around a specific set of elements, and reading a deficiency with that structure in mind is usually more productive than jumping straight to the request at the end. In general, a deficiency should include:

  1. The Issue: what specifically FDA found to be missing, unclear, or inadequately supported in the submission.
  2. Basis for Deficiency: why the information already provided does not resolve the question, and what effect the gap has on the marketing authorization decision.
  3. Information Requested: the specific data, analysis, or explanation FDA needs to move forward, along with any alternative approaches FDA considers acceptable.
  4. Response Window: the date by which a complete response is due.

An applicant must read the basis of the deficiency as the correspondence details it and formulate further action and response from the specific questions asked. 

Timing and Response Windows by Application Type 

A formal AI Request creates a hard deadline. Under 21 CFR 807.87(m), FDA may consider a 510(k) withdrawn if requested information is not submitted within 30 days. Fortunately, current FDA policy extends that window considerably: applicants get up to 180 calendar days from the date of the AI Request to submit a complete response without a written extension request. No extension beyond that period is available.  

For 510(k)s and De Novo requests, the FDA requires a complete response within 180 calendar days that addresses all deficiencies noted. For both, the consequence of missing the 180-day window is that the submission will be considered withdrawn and deleted from FDA’s review system, and the applicant will need to file a new 510(k) to restart the process. 

PMAs operate under a separate regulatory framework: if an applicant fails to respond in writing to an FDA request for an amendment within 180 days, the FDA may consider the PMA voluntarily withdrawn. 

Note that time spent on hold does not count against FDA's own review clock. Under the current MDUFA framework, 510(k)s carry a 90 FDA-day review goal, and the days a submission sits on hold waiting on the applicant's response are excluded from that count.  

The Least Burdensome Principle and its Implications For FDA Interaction

Deficiency requests for device submissions are governed by the FDA’s “least burdensome” principle. “Least burdensome” is defined as the minimum amount of information necessary to resolve a specific regulatory question, through the most efficient approach available, at the appropriate point in review. This means that the FDA should, in addressing deficiencies with an applicant, refrain from asking for extraneous information irrelevant to the applicant device, while maintaining the stringent evidentiary bar for clearance or approval.

Deficiencies generally fall into major and minor deficiency categories; the distinction between these dictates how an applicant should prioritize response. According to the FDA's guidance, Developing and Responding to Deficiencies in Accordance with the Least Burdensome Provisions, they are defined as follows:

  • Major deficiencies are those based on least burdensome principles that, if not resolved, will preclude a favorable marketing authorization decision. 
  • Minor deficiencies can be resolved in a straightforward manner, but still need to be addressed to meet regulatory requirements. They should be resolved, if possible, through interactive review. 

Thus, the applicant should note that a deficiency letter is not an invitation to resubmit the application, and it is not a signal to hand over every document the company has on the topic. The strongest response is well-organized and answers the specific regulatory question(s) the FDA asks. 

Building the Response

For each deficiency, the applicant is generally choosing among the same three paths specified by FDA guidance:

  1. Provide the specific information or data requested;
  2. Explain why the identified issue does not, in fact, affect the marketing authorization decision; or
  3. Present alternative information, along with an explanation of why that information adequately addresses the concern.

It is important to identify what the deficiency requires in response quickly, because the timelines on necessary actions may differ. For example, resolving a deficiency through providing context for data or device specification may be as simple as document review and submission. However, more common with major deficiencies, additional testing may be required, which means significantly more time and incur cost. 

Clarity and targeted information provision are key when building a response. Restate or clearly reference each deficiency in your answer. Answer only what was asked. Breadth of information is not equivalent to a stronger response. Review the package as one submission to ensure descriptions, specifications, software versions, and conclusions match everywhere they appear. A technically correct answer can still raise a new question if it conflicts with something stated elsewhere in the response. 

How to Avoid a Deficiency Letter

The most efficient deficiency response is the one an applicant never has to write. For devices involving novel materials, significant software functions, or non-standard test strategies, early engagement through FDA’s Q-Submission Program can help identify evidentiary gaps before the marketing submission is filed, when there is still time to generate the missing data on a normal timeline. Pre-Submission feedback cannot guarantee a deficiency-free review, but for many applicants, it is a far easier place to resolve open questions than in substantive review, against a 180-day clock.

How Accorto Can Help

Accorto Regulatory Solutions supports device manufacturers across the submission lifecycle, from Pre-Submission strategy through 510(k), De Novo, and PMA preparation and deficiency-response management. Our team can help triage an incoming deficiency letter, identify what requires new testing versus existing evidence, and build a response that answers FDA’s questions completely and on schedule. Learn more on our Regulatory Submissions Solutions page.

austin 1

Austin Tyner

Document Specialist

Austin Tyner is Accorto’s document specialist, supporting the quality and publishing team with document onboarding, formatting, and FDA submission readiness. She holds a BA in English and Comparative Literature, with a minor in Public Policy, from the University of North Carolina at Chapel Hill.

neel v2

Neel Parmar

Regulatory Specialist I

A member of Accorto’s Regulatory Writing Team, Neel Parmar, recently graduated from Northeastern University with a Master of Science in Regulatory Affairs. His academic foundation was enriched by specialized coursework in regulatory compliance, strategic product lifecycle management, and advanced regulatory writing, preparing him to excel in complex regulatory documentation environments. Before his role at Accorto, Neel cultivated valuable experience as part of the team at Interscope Inc., contributing to the transition of a Class IIA medical device from MDD to MDR compliance. His role involved supporting the creation of critical regulatory documents and participating in comprehensive gap analyses to ensure alignment with MDR requirements.